Abstract
Performance characteristics sufficient to provide physical data base specific to the Siemen's Stabilipan 2 orthovoltage x-ray therapy machine are presented. Operating conditions covering the working range of the unit from 100 to 300 kVp are selected. Beam quality, output, the central axis depth dose, relative output factors, field flatness, uniformity index, and filtration characteristics of the beams are studied. Selected results are reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.