Abstract

The COBAS Amplicor CMV Monitor test (Roche Diagnostics), an automated polymerase chain reaction (PCR) assay for the quantification of cytomegalovirus (CMV) DNA in plasma samples, was evaluated in a routine diagnostic laboratory. Using cell culture-derived CMV and CMV-negative human plasma, the linear detection range of the assay as well as its intra-and inter-assay variabilities were assessed. The study design allowed distinguishing variations in results related to amplification and detection from those caused by differences in the efficiency of DNA extraction. The assay was able to identify the majority of samples correctly as positive with CMV DNA concentrations above the limit of detection. However, the reported values were often twofold or more different from the (theoretical) input, which could be explained partly by inefficient DNA extraction. The following values were computed for the coefficients of determination R 2: inter-assay variability excluding DNA extraction, R 2=0.982; including DNA extraction, R 2=0.977; intra-assay variability excluding DNA extraction, R 2=0.992; including DNA extraction, R 2=0.992. On balance, the test has acceptable within-run and between-run reproducibility. It therefore allows the comparison of results obtained at different time-points as well as in different laboratories, e.g. in multi-centre studies.