Abstract
Although the use of implantable cardioverter defibrillator (ICD) therapy has increased over the last decades, the reporting of ICD lead performance is inadequate. So far, there is neither a European nor worldwide registry on ICD leads. The published long-term results from national or multicentre registries encompass relatively small patient cohorts. Nevertheless, the failure of ICD leads may have substantial clinical consequences, including failure to sense, failure to pace, failure to defibrillate, inappropriate shocks, and even death of the patient. The reported ICD lead survival varies significantly between studies: 91-99% at 2 years, 85-95% at 5 years, and 60-72% at 8 years. Thus, the true incidence of lead malfunction cannot be defined as outlined in the present review. One current initiative of the European Heart Rhythm Association is to initiate and develop a Europe-wide registry to monitor, over a prolonged follow-up period, the performance of ICDs and ICD leads.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.