Abstract

To evaluate the functional and microbiological safety of a single-use laparoscopic vessel-sealing device in a multiuse setting. Ex vivo study. Twelve 5-mm LigaSure Maryland jaw devices. Handsets underwent repeated test cycles until failure. The handset packaging was opened, and handsets were agitated in phosphate-buffered saline (PBS). The PBS was removed, centrifuged, and submitted for culture. Canine ovariectomy was simulated on cadaveric tissue, after which vascular seal quality was evaluated by pressure testing of sealed porcine carotid arteries. The handsets were cleaned and sterilized with hydrogen peroxide gas and repackaged. Mean ± SD cycles to failure was 7.7 ± 2.8, with a minimum of 4 and a maximum of 12 use cycles achieved. Eleven of the 12 handsets failed by failure of handset activation after depression of the activation trigger. Only 1 handset failed to hold an adequate vascular seal under 300 mm Hg of pressure. No handset exhibited positive bacterial culture at any cycle. The sterilization method used in this study resulted in an excellent microbiological safety profile. Most of the handsets failed by activation button failure. Under the conditions of this study, hydrogen peroxide sterilization achieved microbiological safety. Handsets can successfully be reused until activation button failure without negative effects on the vascular seal or increased risk of infection to the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.