Abstract
The Belle II Aerogel Ring Imaging Cherenkov (ARICH) counter uses the angular distribution of Cherenkov photons emitted in silica aerogel to discriminate between charged pions and kaons. The Hybrid Avalanche Photo Detector (HAPD) is used as the photo-sensor, and is clearly the most critical component of the detector. HAPDs were installed into ARICH in July 2017, and ARICH was installed in the Belle II spectrometer in the end of 2017. During the Belle II beam commissioning in spring 2018, the performance of HAPDs was evaluated through measurements of the offset value, noise level and pulse height. Long term stability of the signal-to-noise ratio and the fraction of dead channel was also monitored. The signal-to-noise ratio exceeded 6, and the fraction of dead channels was less than 1%. The results of the performance evaluation and of the commissioning showed that the ARICH counter fulfills the requirements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.