Abstract

This paper describes the work on the development of disposable micro sample cell for optical spectroscopy. The disposable flow cell is an attractive approach to ensure cleanliness of the sample container and to avoid contamination between samples and residual detergents. The main objective of this work is to develop a disposable flow cell made of Poly-(methyl methacrylate) (PMMA) with polymer optical fiber (POF) connections that is easy to install and has a performance measurement that meets ISO standards. The research involved three stages namely fiber optic preparation, design of the flow cell and analysis of the flow cell design. Chemometrics methods, internal quality control standards, calibration and performance characterization instruments are used for the analysis. The POF fiber performance is described by a linear calibration graph. The Shewhart chart for uncertainty analysis shows no data out of the chart, with mean value meeting the ISO 8258 standard. Comparisons of the calibration to other disposable sample cells show better results in linearity. Chemometrics analysis specified the reading data to be within the warning line in accordance with ISO 8258. Validation of the mathematical model is acceptable as none exceed 95% for the F-test. Average precision and sensitivity are 0.9 and 0.3, respectively. Limit of detection and limit of quantification are 0.07 and 0.25, respectively. Based on the results, the design of a disposable flow cell with POF fiber meets the standards of analysis for sample containers used in a spectrometer.

Highlights

  • Nitrites are chemical compounds that can cause environmental pollution [1] and have adverse effects on human health [2] besides imposing health risks on marine aquaculture [3].Natural ground water resources are contaminated by this chemical compound [4], especially in ex-landfill sites [5] when microorganisms break down the nitrogen-containing organic compounds from animal manure, sewage waste, and plants [6]

  • RELATED WORK Various studies involving flow cells used as optical sensors in spectroscopy systems can be found in literature

  • The core of the polymer optical fiber (POF) is made of Poly-(methyl methacrylate) (PMMA), which coated by fluorinated polymer and enclosed by polyethylene jacket

Read more

Summary

INTRODUCTION

Nitrites are chemical compounds that can cause environmental pollution [1] and have adverse effects on human health [2] besides imposing health risks on marine aquaculture [3]. S. Abubakar et al.: Performance Analysis of Disposable Flow Cell for Nitrite Detection for quality monitoring process, such as realtime water quality monitoring [11]. Disposable technology [12] in biopharmaceutical industry [13] manufacturing processes is significant such that it can increase efficiency by reducing capital costs [14], improve plant flexibility, reduce start-up time, and eliminate both non-value added process steps and the risk of crosscontamination It reduces labor costs in terms of liquid waste processing and on-site quality and validation requirements [15].

RELATED WORK
POF PERFORMANCE ANALYSIS
COMPARISON OF CALIBRATION
CHEMOMETRICS ANALYSIS
VALIDATION OF THE MATHEMATICAL MODEL
ACCURACY
SENSITIVITY
Findings
VIII. CONCLUSION
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call