Abstract

Nonlinear spatial encoding fields for magnetic resonance imaging (MRI) hold great promise to improve on the linear gradient approaches by, for example, enabling reduced imaging times. Imaging schemes that employ general nonlinear encoding fields are difficult to analyze using traditional measures. In particular, the resolution is spatially varying, characterized by a position-dependent point spread function (PSF). Likewise, the use of nonlinear encoding fields creates an additional spatial dependence on the signal-to-noise ratio (SNR). Although the two properties of resolution and SNR are linked, in this work we focus on the latter. To this end, we examine the pixel variance, which requires a computation that is often not feasible for nonlinear encoding schemes. This paper presents a general formulation for the performance analysis of imaging schemes using arbitrary encoding fields. The analysis leads to the derivation of a practical and computationally efficient performance metric, which is demonstrated through simulation examples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.