Abstract

PurposeTo evaluate the accuracy and technical success of positioning a radiofrequency ablation (RFA) electrode in osteoid osteomas by use of a new real-time needle guidance technology combining cone-beam computed tomography (CT) and fluoroscopy.Materials and MethodsPercutaneous RFA of osteoid osteomas was performed in five patients (median age 18 years), under general anesthesia, with the use of cone-beam CT and fluoroscopic guidance for electrode positioning. The outcome parameters were technical success, meaning correct needle placement in the nidus; accuracy defined as the deviation (in mm) from the center of the nidus; and clinical outcome at follow-up.ResultsIn all five cases, positioning was possible within 3 mm of the determined target location (median nidus size 6.8 mm; range 5–10.2 mm). All procedures were technically successful. All patients were free of pain at clinical follow-up. No complications were observed.ConclusionReal-time fluoroscopy needle guidance based on cone-beam CT is a useful tool to accurately position radiofrequency needles for minimally invasive treatment of osteoid osteomas.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.