Abstract
Percutaneous transluminal laser angioplasty with a pulsed Nd:YAG laser (1,064 nm wavelength, 100 jig pulse duration, up to 0.4 J per pulse, 10 Hz repetition rate) coupled to optical fibers with sapphire tips of 1.8, 2.2 and 2.9 mm diameter was performed in 17 chronic occlusions of iliac, femoral, popliteal and fibular arteries in 15 patients. Clinically the procedure was successful in 14 cases. The probe formed a primary channel of at least 2.0 mm width which was further dilated by conventional balloon catheter. Ankle/brachial systolic pressure index (ABPI) increased from 0.46 0.13 to 0.84 0.21 after the procedure. Two failures were due to an extensive dissection which occurred after balloon angioplasty and resulted in an early reocclusion. In the remaining case a balloon catheter could not be introduced through the 30 cm long primary channel which reoccluded shortly after the recanalization. The initial patency rate was 82 % and in 5 patients followed for more then 6 months the ABPI showed only a minor decrease. These first results are encouraging and it is likely that this method could become an important alternative to conventional balloon angioplasty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.