Abstract

Four months after the successful implantation of a closure device in a 67-year-old woman with an atrial septal defect, the device was found to be impacted in the abdominal aorta. Severe bouts of only partially containable coughing episodes had occurred some days after implantation. Despite unknown endothelialization status and extent of adhesion to the aorta, percutaneous retrieval was favored over conventional surgical retrieval. In conclusion, this case confirms that bouts of elevated intrathoracic pressure can lead to device dislocation and that it may be possible to successfully retrieve a device percutaneously, even months after implantation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.