Abstract

The risk of electromagnetic interference between microwave (MW) ablation and cardiac implantable electronic devices (CIEDs), ie, pacemakers and defibrillators, has not been fully evaluated. Fourteen MW ablations (kidney, n = 8; liver, n = 5; lung, n = 1) were performed in 13 patients with CIEDs in normal operating mode. Electrocardiography tracings, cardiovascular complications, tumor size, tumor-to-CIED distance, and tumor-to-device lead distance were recorded. Mean tumor size was 2.9 cm, mean tumor-to-CIED distance was 26.4 cm (range, 9–30 cm), and mean tumor-to-lead distance was 12.1 cm (range, 3.5–20 cm). No device-based cardiovascular complications or class C or higher complications per Society of Interventional Radiology criteria were identified. MW ablation appears to be safe in select patients with CIEDs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.