Abstract
The objective of this study was to determine if the design enhancements incorporated into the new Provox Vega Indwelling Voice prostheses result in any positive benefits in vivo. Using a randomized, crossover study design, 31 participants using tracheoesophageal speech postlaryngectomy completed a 3-week trial of the Provox Vega and a comparator device. Main outcome measures included patient perceptions of vocal effort and quality using each device, and perceptual judgments of voice quality produced. The majority of patients (72%) indicated the Provox Vega gave them overall better voice quality and 52% felt they required less effort to phonate. Voice samples produced with the Provox Vega were also perceived by listeners to be significantly (p < .05) less strained, easier to understand, produced with less effort, and the better speech sample overall. Results support that the aerodynamic improvements incorporated in the design of the new Provox Vega facilitate enhanced voice and speech qualities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.