Energy resolving performance of spectral CT systems is influenced by the accuracy of the detector's energy calibration. Global energy calibration maps a given threshold to the average energy response of all pixels of the detector. Variations arising from CMOS manufacturing processes and properties of the sensor cause different pixels to respond differently to photons of the same energy. Threshold dispersion adversely affects spectral imaging by degrading energy resolution, which contributes to blurring of the energy information. In this paper, we present a technique for per-pixel energy calibration of photon-counting x-ray detectors (PCXDs) that quantifies the energy response of individual pixels relative to the average response. This technique takes advantage of the measurements made by an equalized chip. It uses a known global energy map to quantify the effect of threshold dispersion on the energy response of the detector pixels across an energy range of interest. The proposed technique was assessed using a MARS scanner with an equalized Medipix3RX chip flip-bonded to 2 mm thick CdTe semiconductor crystal at a pitch of 110 μ m. Measurements were made of characteristic x-rays of a molybdenum foil. Results were compared between the case that the global calibration was used on its own and the case of using it in conjunction with our per-pixel calibration technique. The proposed technique quantified up to 1.87 keV error in energy response of 100 pixels of a selected region of interest (ROI). It made an improvement of 28.3% in average FWHM. The additional information provided by this per-pixel calibration technique can be used to improve spectral reconstruction.

Highlights

  • This content has been downloaded from IOPscience

  • We present a technique for per-pixel energy calibration of photon-counting x-ray detectors (PCXDs) that quantifies the energy response of individual pixels relative to the average response

  • The analysis presented includes experimental results of the proposed per-pixel calibration technique that was used in conjunction with global energy calibration using the kVp method [7]

Read more

Summary

Introduction

This content has been downloaded from IOPscience. Please scroll down to see the full text. 2017 JINST 12 C03085 (http://iopscience.iop.org/1748-0221/12/03/C03085) View the table of contents for this issue, or go to the journal homepage for more. 2017 JINST 12 C03085 (http://iopscience.iop.org/1748-0221/12/03/C03085) View the table of contents for this issue, or go to the journal homepage for more. We present a technique for per-pixel energy calibration of photon-counting x-ray detectors (PCXDs) that quantifies the energy response of individual pixels relative to the average response. This technique takes advantage of the measurements made by an equalized chip. It uses a known global energy map to quantify the effect of threshold dispersion on the energy response of the detector pixels across an energy range of interest

Objectives
Methods
Findings
Discussion
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call