Abstract
This article discusses the performance of Vibrating Sample Magnetometer (VSM) Oxford 1.2 H in characterizing the magnetic samples in liquid form. Samples of water-based ferrofluid placed in a capillary tube pipe. Measurement of samples with varied position, volume and concentration as well as repeated measurements are carried out to evaluate the performance. From these series of measurements, it brings that optimal volume range with measurement results that can still be accounted for is 10 µl, with samples being in equilibrium position within the pick-up coil. The smallest sample concentration which still can be observed quantitatively is 0.332 mg/mL. Repeatability test gives the value of the magnetic moment deviation of ~ 4%. The accuracy of the sample preparation and positioning of the samples is analyzed being a major factor which causes this measurement error. In general it can be concluded that the liquid sample measurement will give better results when using reverse pipetting mode when putting the liquid samples within capillary tube using a micropipette. The measurement should be done as soon as possible after preparation. DOI: https://doi.org/10.17977/um024v2i22017p039
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: JPSE (Journal of Physical Science and Engineering)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.