Abstract
This study aims to determine the effect of using the high kV technique on chest radiographic examination on Entrance Surface Air Kerma (ESAK), and Dose Area Product (DAP) received by patients. This is one of the optimization efforts in radiation protection and safety in ionizing radiation. This study used retrospective data from chest radiographic images generated by the Siemens Ysio Max radiograph at the Radiology Installation of RSUP Dr. Kariadi Semarang in 2021. Two voltages were used, 59.8 kV and 108.8 kV, with a total sample of 225 patients. The use of the high kV technique produces a smaller ESAK value, namely with a difference of 60.75%, and a smaller DAP value, namely with a difference of up to 74.92%, compared to using the standard kV technique. These results show a significant difference in the radiation dose produced using the high kV technique with image quality that can still use to diagnose a disease/disorder. Keywords: chest radiography, high kV technique, Entrance Surface Air Kerma, Dose Area Product
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.