Abstract

Two algorithms for solving misalignment issues in penalized PET/CT reconstruction using anatomical priors are proposed. Both approaches are based on a recently published joint motion estimation and image reconstruction method. The first approach deforms the anatomical image to align it with the functional one while the second approach deforms both images to align them with the measured data. Our current implementation alternates between alignment estimation and image reconstruction. We have chosen parallel level sets (PLSs) as a representative anatomical penalty, incorporating a spatially variant penalty strength. The performance was evaluated using simulated nontime-of-flight data generated with an XCAT phantom in the thorax region. We used the attenuation map in the anatomical prior. The results demonstrated that both methods can estimate the misalignment and deform the anatomical image accordingly. However, the performance of the first approach depends highly on the workflow of the alternating process. The second approach shows a faster convergence rate to the correct alignment and is less sensitive to the workflow. The presence of anatomical information improves the convergence rate of misalignment estimation for the second approach but slow it down for the first approach. Both approaches show improved performance in misalignment estimation as the data noise level decreases.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call