Abstract

Objective To determine the accuracy of the MTI Photoscreener in detecting strabismus and refractive errors in children. Participants One hundred children underwent MTI photoscreening followed by complete ophthalmologic examination. Six observers graded the photographs for strabismus, according to the location of the corneal light reflexes, and for refractive error, according to the size and location of the light crescent. Results The sensitivity of the MTI Photoscreener in detecting any amblyogenic factor was 80% to 91%, with a specificity of 20% to 67%. The sensitivity and specificity for particular amblyogenic factors varied widely among observers. The ranges were as follows: strabismus, sensitivity = 23% to 50%, specificity = 76% to 96%; myopia, sensitivity = 89%, specificity = 48% to 76%; hyperopia, sensitivity = 20% to 80%, specificity = 88% to 96%; and astigmatism, sensitivity = 46% to 77%, specificity = 79% to 89%. Conclusions These results suggest caution in relying on photoscreening to detect strabismus and refractive errors in children.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.