Abstract

The Impella CP is a percutaneously inserted temporary left ventricular assist device used in clinical practice and in translational research into cardiogenic shock, perioperative cardiac surgery, acute cardiac failure and mechanical circulatory support. Fluoroscopic guidance is usually used for insertion of an Impella, thus limiting insertion to within catheterization laboratories. Transthoracic, transoesophageal and intracardiac echocardiography have been reported to guide Impella CP implantation with identified specific limitations stemming from the surgical, anatomical and equipment factors. We conducted translational prospective descriptive feasibility investigation as a part of two other hemodynamic Impella studies. It showed the successful application of epicardial echocardiographic scanning for implantation of Impella CP devices in ovine models, from which details of the technique and identified pitfalls are described with practical solutions for future investigators and clinicians. Many described findings are relevant to any other echocardiographic techniques when adequate imaging of the Impella and relevant anatomical structures is achievable.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.