Abstract
The behavior of voltammograms in square wave voltammetry in the presence of a significant ohmic potential drop for reversible and irreversible simple electrode reactions has been investigated. It is shown that the effects of ohmic potential drop can be drastically reduced by choosing adequately the square wave amplitude. Zone diagrams are drawn to define the conditions for which peak currents and peak widths at half-height can be expressed by simple formulas without considering the occurrence of ohmic potential drop. It follows that square wave voltammetry can be properly implemented in low-conductive media or in molten salt media where high concentrations of depolarizer are used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.