Abstract
This paper describes the measurement of the peak and recovery heat flux densities ( q P, q R) in a bath of subcooled superfluid helium. A RhFe wire ( d = 51 μm, L = 34.8 mm) was horizontally immersed in a He Ilp cryostat, which served as the sample, the heater and the temperature sensor. For the steady state measurement a programmed voltage method with a triangular waveform of 20 s was applied. The measured q P and q R values cover the temperature range 1.81–2.1K at a pressure of 0.1 MPa. On the basis of the heat and mass transfer through phase interfaces, theoretical relations (without geometrical parameters) of q P and q R are derived, which have very wide application range due to the relative critical heat flux density method used for the calculations. The theoretical results fit the experimental results well for both q P and q R.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.