Abstract

The revision of an hip prosthesis can have diffrent reasons. One frequent cause, especialley after implantation of a conventional cup, is the so called stress-shielding ef- fect which can lead to a migration or loosening. Patient- specific hip cups can be used to counteract this. However, individual hip cups are only implanted for the treatment of great deformations or tumours because of the cost-intensive manufacturing. Within this project a patient-specific hip cup prosthesis has to be developed and manufactured. Be- sides the numerical design by means of a coupling between multi-body simulation (MBS) and finite element method (FEM), an inovative concept for the production of patient- individual hip prosthesis out of titanium sheets is introduced in this study. Patient-individual hip cups, bone remodelling, reverse en- gineering, process planning, sheet metal forming

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.