Abstract

» Universal safety guidelines for the use of a knee-spanning external fixator in magnetic resonance imaging (MRI) are unlikely to be established considering the high variability in device construct configurations.» Per the US Food and Drug Administration, manufacturers are to provide parameters for safe MRI scanning for "MR Conditional" devices; however, such labeling may be limited in detail. Physicians should reference manufacturer labels as a starting point while making an educated clinical decision.» Scanning of a knee-spanning external fixator inside the MR bore has been safely demonstrated in previous studies, although with small sample sizes.» When considering MRI in a patient treated with a knee-spanning external fixator, physicians should use all available resources and coordinate with their medical team to make a clinically reasonable decision contrasting patient benefit vs. potential harm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.