Abstract
This study aims at accurate quantification of x-ray exposure and effective dose to the patient in abdominal arteriography. Using an automatic monitoring system, all relevant exposure parameters were determined during 172 abdominal arteriographies. Common projections were extracted for a `normal' reference group of procedures and used in Monte Carlo calculations of dose-area product to organ dose conversion coefficients. Dose-area product, organ doses and effective dose were quantified for intravenous and intra-arterial procedures. The large data sets describing exposure could be condensed to a set of 28 common views. New coefficients to convert dose-area product to organ equivalent dose and effective dose were calculated for nine views contributing approximately 80% to the total dose-area product. The average dose-area product was in intravenous procedures and in intra-arterial procedures. The corresponding average effective doses to the patient were 4 mSv and 6 mSv respectively (range 2-12 mSv, actual value depending on procedure type and gender). It is concluded that automatic monitoring of x-ray exposure parameters, complemented by the calculation of Monte Carlo organ dose conversion coefficients, is a feasible and promising approach to accurate dosimetry of complex arteriographic procedures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.