Abstract

The purpose of this prospective study at 73 facilities in 18 countries in Africa, Asia and Eastern Europe was to investigate if the CT doses to adult patients in developing countries are higher than international standards. The dose assessment was performed in terms of weighted computed tomography dose index (CTDIw) and dose length product (DLP) for chest, chest (high resolution), lumbar spine, abdomen and pelvis CT examinations using standard methods. Except in one case, the mean CTDIw values were below diagnostic reference level (DRL) while for DLP, 17 % of situations were above DRLs. The resulting CT images were of adequate quality for diagnosis. The CTDIw and DLP data presented herein are largely similar to those from two recent national surveys. The study has shown a stronger need to create awareness and training of radiology personnel as well as monitoring of radiation doses in many developing countries so as to conform to the ALARA principle.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.