Abstract

We analyzed 18 explanted 91Z anterior chamber lenses by light and scanning electron microscopy. Intermittent touch of the lens loops to the posterior corneal surface and the anterior chamber angle may have occurred. Erosion of the polypropylene loops into the anterior chamber angle recess and into the iris tissue was also observed. Fibrous tissue, uveal tissue, and inflammatory debris were noted on the loops, forming dense synechias at the points of contact with the angle recess. In some lenses the edges of the optics were sharp. Other significant manufacturing defects were rarely seen, and there was no evidence of degradation of polypropylene loops. The problems regarding surgical removal of this lens are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.