Abstract

The purpose of this study was to evaluate the safety and effectiveness of the SeptRx patent foramen ovale (PFO) closure device (SeptRx, Inc., Fremont, California). A PFO is a relatively common remnant of the fetal circulation that can be associated with cryptogenic stroke, transient ischemic attack, migraine, or decompression sickness. Percutaneous PFO closure with different devices has been performed for many years. However, most of the common devices leave a relatively large amount of material in the left and right atria. The SeptRx PFO device (SeptRx, Inc.) is the first PFO closure device designed to fit directly into the pocket of the PFO. From July 2006 to May 2007, 13 patients between 18 and 65 years of age with a history of cryptogenic stroke or transient ischemic attack were included into this first-in-man trial. All patients received 100 mg aspirin and 75 mg clopidogrel for 6 months. Follow-up was done at 1 and 6 months after procedure with transesophageal echocardiography and transcranial Doppler. In 11 of 13 patients, PFO closure with the SeptRx device was successfully performed. In 2 patients, PFO closure with this device was not possible due to the anatomy of the PFO; 1 device was retrieved before release, and the other was recaptured with a snare. After 30 days, 6 of the 11 PFOs were closed; after 6 months, all were closed. No adverse events occurred. The SeptRx PFO closure device appears to be safe and effective. The advantage of this occluder is that there is only minimal foreign material on the left and right sides of the interatrial septum. This is the first such "in-tunnel" PFO closure device.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call