Abstract

In the setting of an Ebstein's anomaly, a right-to-left shunt via a patent foramen ovale (PFO) may be the cause of a significant cyanosis. In these patients, the PFO closure is able to improve the arterial saturation. Furthermore, a partial closure is often recommended to obtain a reduction in the PFO right-left shunt without a significant impact on the right chamber hemodynamics. However, in some cases, a complex PFO anatomy may prevent an effective device closure. We describe the case of a patient with Ebstein's anomaly and a significant cyanosis due to a right-to-left shunt via the PFO wherein an effective partial PFO closure with a suture-mediated NobleStitch EL system was achieved in a setting of a complex PFO anatomy (large aneurismatic septum, long stiff tunnel, and hypertrophic septum secundum) after a failed device-closure attempt.