Abstract

Patellar instability following total knee replacement although infrequent can cause disabling symptoms. We report a case of symptomatic patellar instability caused by insertion of left femoral component into a right knee. We highlight that all members of the surgical and theatre team need to be familiar both with the implant equipment and of any recent design modifications, particularly if they have been used to using one system for a long time, in order to prevent surgical errors. In addition this possibility should be born in mind by surgeons looking for causes of post-operative patellar instability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.