Abstract

Planar patch-clamp has revolutionized ion-channel measurement by eliminating laborious manipulation from the traditional micropipette approach and enabling high throughput. However, low yield in gigaseal formation and/or relatively high cost due to microfabricated processes are two main drawbacks. This paper presents patch clamping on glass substrate-an economical solution without sacrificing gigaseal yield rate. Two-stage CO(2) laser drilling methodology was used to generate an hourglass, funnel-like aperture of a specified diameter with smooth and debris-free surfaces on 150 microm borosilicate cover glass. For 1-3 microm apertures as patch-clamp chips, seal resistance was tested on human embryonic kidney, Chinese hamster ovary, and Jurkat T lymphoma cells with a gigaseal success rate of 62.5%, 43.6% and 66.7% respectively. Results also demonstrated both whole-cell and single channel recording on endogenously expressed ion channels to confirm the capability of different patch configurations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.