Abstract

ObjectiveThis study aimed to compare the accuracy of the endotracheal tube (ETT) cuff pressure of the manual palpation (MP), passive release (PR), and minimum occlusive volume (MOV) techniques.MethodsThis study is a true experiment with simple randomisation. The subjects of this study were 105 patients divided into 3 groups: MP group (n=35), PR group (n=35), and MOV group (n=35). After intubation, ETT cuff inflation was performed using 3 different techniques. The ETT cuff pressure was recorded using a manometer. The data were analysed using the chi-square test, Kruskal-Wallis test, and Mann-Whitney test in the SPSS 20 software.ResultsThe mean ETT cuff pressure was 60.2±28.8 cmH2O in the MP group, 30.4±5.5 cmH2O in the PR group, and 25.8±9.6 cmH2O in the MOV group (p=0.000). The PR group had the highest pressure accuracy (77%) (p=0.000).ConclusionThe PR technique had the highest accuracy and can be used as an alternative ETT cuff inflation technique in the absence of a manometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.