Abstract

A passive neutron dosemeter was designed to be used in mixed radiation fields. The design was carried out using Monte Carlo method. The dosemeter model was a 25.4 cmdiameter polyethylene sphere with a thermoluminescent dosemeter, TLD600, located at the sphere center. This model was irradiated with 50 monoenergetic neutron sources with energies from 10−8 to 20 MeV. A 506.71 cm2‐area disk was used to model the source term whose center was located at 100 cm from polyethylene sphere's center. The dosemeter response was compared with the responses of SNOOPY, Harwell 95/0075 and PNR‐4. With these responses it was calculated the dosemeter responses for 252Cf, 252Cf/D2O and 239PuBe neutron sources. The passive dosemeter relative response has the same shape of SNOOPY, Harwell 95/0075 and PNR‐4 dosemeters. Due to the type of thermal neutron detector used in the passive dosemeter the absolute response per unit fluence, is lower than the absolute response of SNOOPY, Harwell 95/0075 and PNR‐4 dosemeters. However the passive dosemeter response in function of the average neutron energy of the 252Cf, 252Cf/D2O and 239PuBe neutron energy was more linear.