Abstract

In this paper a hybrid system able to provide focused microwave radiometry and deep brain hyperthermia is experimentally tested. The system's main module is an ellipsoidal conductive wall cavity which acts as a beam former, focusing the electromagnetic energy on the medium of interest. The system's microwave radiometry component has extensively been studied theoretically and experimentally in the past few years with promising results. In this work, further investigation concerning the improvement of the hybrid system's focusing properties is conducted. Specifically, microwave radiometry and hyperthermia experiments are performed using water phantoms surrounded by dielectric layers used as matching material to enhance detection/penetration depth and spatial resolution. The results showed that the dielectric material reduces the reflected electromagnetic energy on the air–phantom interface, resulting in improved temperature resolution and higher detection or penetration of the energy when microwave radiometry and hyperthermia are applied respectively.