Abstract

Wireless passive neural recording systems integrate sensory electrophysiological interfaces with a backscattering-based telemetry system. Despite the circuit simplicity and miniaturization with this topology, the high electrode-tissue impedance creates a major barrier to achieving high signal sensitivity and low telemetry power. In this paper, buffered impedance is utilized to address this limitation. The resulting passive telemetry-based wireless neural recording is implemented with thin flexible packages. Thus, the paper reports neural recording implants and integrator systems with three improved features: (1) passive high impedance matching with a simple buffer circuit, (2) a bypass capacitor to route the high frequency and improve mixer performance, and (3) system packaging with an integrated, flexible, biocompatible patch to capture the neural signal. The patch consists of a U-slot dual-band patch antenna that receives the transmitted power from the interrogator and backscatters the modulated carrier power at a different frequency. When the incoming power was 5-10 dBm, the neurosensor could communicate with the interrogator at a maximum distance of 5 cm. A biosignal as low as 80 µV peak was detected at the receiver.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call