Abstract
We present passive flow-rate regulators using parallel membrane valves rectifying dynamic inlet pressure. The parallel membrane valves in the passive flow-rate regulator are designed to deflect and adjust the flow resistance according to the inlet pressure, thus maintaining a constant flow-rate independent of the inlet pressure variation. The prototype was fabricated by PolyDiMethylSiloxane (PDMS) micromolding process. In the experimental study, the dynamic inlet pressure in the range of 20 and 50 kPa was applied at the varying frequency of 1∼80 Hz and the outlet flow-rate was measured. The prototype maintained as the constant flow-rate of 5.82±0.29 µl/s at the dynamic inlet pressure. The present passive flow-rate regulator has shown the strong potential for the precise sample delivery in integrated microfluidic systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.