Abstract

To investigate whether a passive electrode effect decreases defibrillation threshold (DFT) in multi-filament middle cardiac vein (MCV) defibrillation. Twelve pigs underwent active housing (AH) insertion, with defibrillation coils placed transvenously in right ventricular apex and superior vena cava. The MCV was cannulated, and 1.12F, 50 mm coil electrodes (Ela Medical SA, France) were deployed in its right and left branches. Lead placement was possible in 11 of 12 animals. DFT (J, mean +/- SD) was determined by three-reversal binary search and compared between the MCV monofilament (single filament deployed) and the AH (25.9 +/- 10.9) and the MCV mono + passive filaments (both filaments deployed, one connected) and the AH (19.9 +/- 11.4); 24% DFT reduction P = 0.008. A bystander electrode adjacent to a monofilament electrode in the MCV reduces DFT by 24% when compared with monofilament MCV alone. Microfilament electrodes decrease DFT as auxiliary anode but not as sole anode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.