Abstract
Herein, we present a simple microelectrode preparation methodology consisting in coating a platinum wire or a carbon fiber with a thin insulating Parylene C film (e.g. 1–10μm), to produce SECM probes with a small and constant probe RG (i.e. ratio between the radius of the insulating sheath and the radius of the active electrode area). After exposition of a fresh active electrode area by blade cutting, a disc shaped electrode is obtained thanks to a protective hot mounting wax layer that avoids Parylene C coating deformation and is easily removed with acetone. Stiffness and straightness of the probe can be tuned by modifying the Parylene C coating thickness and the length of the carbon fiber or platinum wire. This simple electrode preparation method is highly reproducible (c.a.>90%). The prepared Parylene C coated microelectrodes were characterized by optical microscopy, cyclic voltammetry, scanning electrochemical microscopy (SECM) approach curves and finally applied to SECM imaging of Pt band structures in contact-less and contact mode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.