Abstract

We introduced frequency-control and temperature-control systems in wavelength-meter-controlled cavity ring-down spectroscopy. The frequency-control system shifted the wavelength of the probe laser from 1393 nm to 696.5 nm where no strong absorption lines of water exist, and therefore, it could avoid measurement errors in laser frequency due to residual moisture in the built-in Fizeau interferometer of a wavelength meter. We verified the hypothesis that the nonuniform reflectivity of the mirror surface contributes to fluctuations in the ring-down time observed with multi-transverse-mode CRDS signals. The fluctuations due to this effect were greatly suppressed by the temperature-control system. Using this system, we could improve the minimum detectable absorption coefficient by three times on average and also improve the experimental standard deviations of the averages by nine times compared with those without the system. We measured near-infrared spectra of residual moisture in dry nitrogen at an approximately 1 nmol/mol (1 ppb) level and performed least-squares fitting of the averaged spectrum. The standard deviation of the residuals of the fitting was 6.6 × 10−12 cm−1, corresponding to a mole fraction of water of 6.3 pmol/mol (6.3 ppt).

Highlights

  • The demand for accurate measurement of trace moisture in high-purity industrial gases at a level of 10 nmol/mol (10 ppb) or lower has been increasing in the semiconductor industry

  • One issue in using a high-finesse cavity is that the frequency of the probe laser must be accurately and precisely controlled to match the resonant frequency of the cavity; Cavity ring-down spectroscopy (CRDS) can perform measurement only when the frequency of a laser matches the resonant frequency of the cavity

  • In wavelength-meter-controlled CRDS, the frequency of the laser is controlled to match a resonant frequency of the cavity on the basis of the frequency measured by a wavelength meter

Read more

Summary

Introduction

The demand for accurate measurement of trace moisture in high-purity industrial gases at a level of 10 nmol/mol (10 ppb) or lower has been increasing in the semiconductor industry. Parts-per-trillion sensitivity for trace-moisture detection using wavelength-meter-controlled cavity ring-down spectroscopy

Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.