Abstract

We characterized the particle size distribution and the analyte transmission efficiency of a liquid chromatography/particle beam/mass spectrometry (LC/PB/MS) system as a function of experimental variations normally used to optimize the LC/PB/MS system. The particle size distribution was evaluated using an electrical differential mobility particle sizer (DMPS) and both the DMPS and the mass spectrometer were used to evaluate transmission. The latter results were correlated to provide evidence related to mechanisms which contribute to poor sample transmission. Addition of ammonium acetate buffer did not increase the aerosol particle mean diameter. However, it did lead to significant increases in caffeine transmission efficiency observed in both the DMPS and the mass spectrometer. Our results were interpreted to suggest a possible electrostatic cause for quantitative anomalies in LC/PB/MS rather than simple mass discrimination in the particle beam momentum separator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.