Abstract

Magnetic resonance temperature imaging can be used to monitor the progress of thermal ablation therapies, increasing treatment efficacy and improving patient safety. High temporal resolution is important when therapies rapidly heat tissue, but many approaches to faster image acquisition compromise image resolution, slice coverage, or phase sensitivity. Partially parallel imaging techniques offer the potential for improved temporal resolution without forcing such concessions. Although these techniques perturb image phase, relative phase changes between dynamically acquired phase-sensitive images, such as those acquired for MR temperature imaging, can be reliably measured through partially parallel imaging techniques using reconstruction filters that remain constant across the series. Partially parallel and non-accelerated phase-difference-sensitive data can be obtained through arrays of surface coils using this method. Average phase differences measured through partially parallel and fully Fourier encoded images are virtually identical, while phase noise increases with g(sqrt)L as in standard partially parallel image acquisitions..

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.