Abstract

In computed tomography (CT), the term "artifact" is applied to any systematic discrepancy between the CT numbers in the reconstructed image and the true attenuation coefficients of the object. A partial ring artifact in cardiac CT has a unique pattern and has not been previously reported in real clinical practice. In this report, we will demonstrate the images and animations of a cardiac CT which is affected by a partial ring artifact due to a broken detector module. The different characteristics of ring artifacts on routine and cardiac CT, and the implications for clinical interpretation will also be described. In summary, when a partial ring artifact is noted incidentally on cardiac CT, we suggest interpreting the images by reviewing reconstructed images from different RR-intervals to avoid unnecessary repeated scans and additional radiation exposure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.