Abstract
The method of reference transfer is applied to the full range partial pressure calibration of residual gas analyzers (RGA). Partial pressures are transferred across the upper orifice of a 10× pressure divider in such a way that a mathematically interrelated series of increasing partial pressures is generated. A gas flow measurement is used to establish the value of the partial pressure when it is in the 10−5 to 10−7 Torr region. For gases not residual to the vacuum system, calibration can begin at pressure levels limited primarily by electronic sensitivity and noise. Independence of RGA peak heights from interference by certain other gases was observed over a wide range of partial pressures of the several gases. Using this same equipment, inert gas fixed leaks were calibrated for just the inert gas components of their effluent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.