Abstract

Controversy still exists on the validity of processing a partial bowl during the collection of shed blood lost through surgery during cell salvaging. The purpose of this study was to assess the quality of red blood cells produced from a partial bowl of autologous suctioned blood using the Haemonetics Cell Saver 5. Suctioned blood was collected from 17 patients undergoing cardiac surgery. A partially filled cell saver bowl was washed with 1500 mL of NaCl. Reservoir and processed blood samples were examined for potassium, leukocytes, hematocrit, platelets, and plasma-free hemoglobin and then compared with 22 previously studied full bowls. Results are summarized in the table below:In conclusion, the Haemonetics Cell Saver 5 can produce a quality product from washing a partial bowl with a better washout of white blood cells compared with a full bowl. However, there is a reduction in red blood cell recovery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.