Abstract

To evaluate parents' ability to accurately assess their child's heart and respiratory rates (RRs) in the context of potential utility for telehealth visits. In this controlled study of 203 child-parent pairs, parents measured their child's heart rate (HR) using 4 methods: palpation, auscultation, and 2 photoplethysmographic smartphone applications. Parents measured RR by inspecting the child and tapping the smartphone application. The gold standards were electrocardiogram for the HR and the child's breaths measured by a health care professional for 60seconds for the RR. We plotted the measurements using a Bland-Altman plot with 95% limits of agreement. Parents underestimated HR by palpation with a calculated bias of -18 beats per minute (bpm) (SD, 19), with limits of agreement ranging from -56 to 19 bpm. Parents overestimated and underestimated HR by auscultation with limits of agreement ranging from -53 to 46 bpm. Smartphone applications did not improve the accuracy of measurements. The accuracy of parental RR measurements was low. For young children, bias was -0.8 breaths per minute (brpm) (SD, 9.8) with limits of agreement from -20 to 19 brpm, and for older children, bias was 0.9 brpm (SD 7.4) with limits of agreement from 6 to 15 brpm. The sensitivity of parental subjective opinion to recognize accelerated RR was 37% (95% CI, 25%-51%). Parents were not able to assess their child's RR or HR accurately. Digital remote assessment of children should not rely on parental measurements of vital signs.