Abstract

Radiofrequency coils are utilized during transmit and receive of MRI signals. Cable traps remove common-mode current from the coaxial cable shield, which helps improve the image quality and reduces risks of burns to the patient. Traditional cable traps use wounded coaxial cables that limit the flexibility in the design process. Floating cable traps were introduced which eliminated any physical connection between the trap and coaxial cable, allowing complete flexibility in design and placement. However, the design process of floating cable traps is iterative and may take several rounds of 3D modeling. This work seeks to optimize the design process through the use of parametric design methodologies. The proposed methodology allows for 3D printing the floating cable trap after inputting the design parameters. The cable trap was able to attenuate currents in the coaxial shields to -48 dB, highlighting its performance and design robustness.