Abstract
The parameters of the ISO-4037 wide spectrum (W-series) X-ray reference fields were simulated using the SpekPy web-based simulation software. One of the most important physical quantities (air Kerma rate) and other radiation quantities (X-ray fluence spectrum and its averaged energy, the first and the second half value layers, as well as the beam homogeneity coefficient) were also deduced. The comparison in radiation quantities between the simulated (obtained from this work) and the ISO-published data was performed to confirm the reliability of SpekPy simulations (the maximum discrepancy between them was less than 4.0%). The air Kerma rate trend as a function of the W-series X-ray reference field was investigated and presented in this work. The values of air Kerma rates can be applied in calibrations of wide-spectrum X-ray measuring devices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.