Abstract
The lifetime of a macromolecular crystal in an X-ray beam is assumed to be limited by the absorbed dose. This dose, expressed in Gray (Gy = J kg(-1)), is a function of a number of parameters: the absorption coefficients of the constituent atoms of the crystal, the number of molecules per asymmetric unit, the beam energy, flux, size and profile, the crystal size, and the total irradiation time. The effects of these variables on the predicted absorbed dose, calculated using the program RADDOSE, are discussed and are illustrated with reference to the irradiation of a selenomethionine protein crystal of unknown structure. The results of RADDOSE can and will in the future be used to inform the data collection procedure as it sets a theoretical upper limit on the total exposure time at a certain X-ray source. However, as illustrated with an example for which the experimental data are compared with prediction, the actual lifetime of a crystal could become shorter in those cases where specific damage breaks down crucial crystal contacts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.