Abstract

We describe a comparative investigation of the parameters of MoAu-bilayer TES bolometers designed for infrared detectors. A set of devices with variations in geometry were fabricated at the NASA/GSFC detector development facility. These detectors have different bilayer aspect ratios (providing differing normal state resistances and current densities), and have varieties of normal metal regions to study the effects of geometry on noise. These normal metal regions are oriented either parallel to or transverse to the direction of current flow, or both. The lowest noise detectors are found to have normal metal regions oriented transversely. For about a dozen different devices, we have measured a large set of parameters by means of a suite of tests. These include complex impedance measurements to derive time constants; IV curves to determine resistance and power; thermal conductance measurements; noise measurements as a function of device resistance; and direct resistance vs. temperature measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.