Abstract

There are two widely used distal humerus fracture (DHF) fixation methods with either orthogonal or parallel double-plate osteosynthesis. However, biomechanical studies have shown inconsistent results on which technique is more effective. We performed a meta-analysis to compare these two fixation methods for adult DHF fixation. We searched the literature for entries discussing the biomechanical testing of orthogonal and parallel fixation techniques for DHFs. We then performed a meta-analysis of the following biomechanical outcome measures: axial/sagittal/coronal/torsional stiffness, load to failure, and torque to failure. Seventeen studies comparing both constructs were included. The parallel configuration exhibited greater mechanical strength with respect to axial stiffness/load to failure, torsional stiffness, and posterior bending load to failure than the orthogonal constructs. Subgroup analysis revealed that parallel constructs also had higher torsional stiffness in supracondylar fractures. This meta-analysis shows that parallel constructs provide greater axial stiffness, axial strength, and torsional stiffness than orthogonal plate for DHF fixation. A subgroup analysis revealed that parallel constructs had better torsional stiffness in supracondylar fracture fixation. IA.