Abstract

We evaluated the new Beckman multichannel analyzers Astra 8 and Astra 4. Both instruments performed the following tests: Na, K, Cl, CO2, urea, glucose, and creatinine in plasma and Na, K, Cl, and creatinine determinations in urine. We tested precision, accuracy, and linearity at the usual concentrations in plasma and urine. We compared the Astra 8 and Astra 4 with continuous-flow (SMA 6) and discrete analyzers such as IL 243, Beckman KLiNa Flame, Beckman System 1, Corning 920 M Chloride Meter, and the Corning Blood-Gas Analyzer 175. Special tests were performed on plasma-to-plasma and urine-to-plasma carryover. Both analyzers are easy to operate and suitable for both emergency and routine use.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call