Abstract

A novel approach to achieve the image restoration is proposed in which each detector’s relative position in the detector array is no longer a necessity. We can identify each detector’s relative location by extracting a certain area from one of the detector’s image and scanning it on other detectors’ images. According to this location, we can generate the point spread functions (PSF) for each detector and perform deconvolution for image restoration. Equipped with this method, the microscope with discretionally designed detector array can be easily constructed without the concern of exact relative locations of detectors. The simulated results and experimental results show the total improvement in resolution with a factor of 1.7 compared to conventional confocal fluorescence microscopy. With the significant enhancement in resolution and easiness for application of this method, this novel method should have potential for a wide range of application in fluorescence microscopy based on parallel detecting.